STRIDE-HC Walkthrough — from blank page to completed threat model

A first-time-user tutorial. We build a complete STRIDE-HC threat model from scratch using a hypothetical legacy infusion pump as the running example. By the end you’ll have a populated stride-hc-template.md file that you can compare against the worked example in examples/infusion-pump.md.

Estimated time: 30 minutes for a first attempt, 10–15 minutes once you’ve done one.

Audience: anyone who can read the paper. No prior threat-modelling experience required; this walkthrough explains what STRIDE-HC asks for and how to answer it for legacy clinical devices.

What you need before starting

Have these facts about the device in front of you:

If you don’t know some of these, that’s fine — fill in “unknown” and treat closing the gap as a Phase 0 deliverable. A STRIDE-HC threat model with explicit “unknown” fields is more useful than one with assumptions baked in.


Step 1 — Determine the device archetype (2 minutes)

STRIDE-HC distinguishes two archetypes because the feasible compensating controls differ substantially.

If the device runs… Archetype Typical examples
Windows / Linux / general-purpose Unix on standard PC-class hardware Archetype 1 PACS workstations, imaging modality consoles (CT, MRI), laboratory analysers, pharmacy automation, radiotherapy planning stations
Embedded RTOS (VxWorks, QNX, ThreadX, FreeRTOS, etc.) or proprietary firmware Archetype 2 Infusion pumps, ventilators, physiological monitors, anaesthesia machines, dialysis controllers

How to decide if you’re not sure:

For our example pump: runs VxWorks 6.9, no user-visible OS, monolithic firmware image — Archetype 2.

The archetype determines which scenario library you draw from:


Step 2 — Copy the template, fill the device profile (5 minutes)

Copy stride-hc-template.md to a new file named for your device. Recommended naming: <class>-<model>-<inventory-tag>.md. For our pump: infusion-pump-empump-volumetric-v32.md.

Open the new file. Section 1 is the device profile table. Every field has italicized inline guidance that you replace with real values. For our pump:

Field Value
Device name and model ExampleMed Volumetric Infusion Pump v3.2
Manufacturer ExampleMed Inc.
Device class (FDA) Class II
UMDNS code 16495
GMDN code 35143
MDS² reference ExampleMed-IP3.2-MDS2-2024
Archetype Archetype 2 (embedded RTOS legacy)
Operating system VxWorks 6.9 (EOL — no further patches)
Networking 100Mbit Ethernet, HL7 v2 over TCP/2575, no TLS
Authentication Hardcoded service-mode credential; no per-user login
Patching Manufacturer no longer issues patches; FDA clearance limits modification
Audit logging None on device; only at upstream nursing station
Physical interfaces RS-232 service port; USB-A (firmware update); Bluetooth (paired display)
Deployment 240 units across ICU and oncology
Clinical use Continuous medication infusion

This becomes the shared context that every later section refers to. Get it right; everything else flows from it.

Tip: the MDS² disclosure form (if the vendor provides one) answers most of these fields directly. If the device pre-dates MDS² or the vendor didn’t issue one, contact vendor support for the equivalent answers — vendors are increasingly under FDA §524B pressure to provide them retroactively.


Step 3 — Work through the six STRIDE-HC categories (15 minutes)

This is the substantive work. For each STRIDE letter, you answer two attacker questions and document two response questions.

The four-block pattern (applies to every category)

### X — Category name (CAW = w)

**Network-attacker scenarios:**
- Scenario 1
- Scenario 2
- ...

**Physical / insider-attacker scenarios:**     ← STRIDE-HC's key addition
- Scenario 1
- Scenario 2
- ...

**Detection methods (Framework V):**
- Method 1
- Method 2
- ...

**Compensating controls (Framework I):**
- Control 1
- Control 2
- ...

**CCD coverage assessment:** <how well do the controls cover the scenarios>

You don’t have to invent scenarios from scratch

Open your archetype’s scenario library and copy the relevant scenarios into each category. The library lists common scenarios per category for legacy clinical devices. Your job is:

  1. Copy the scenarios that apply to your device.
  2. Cut the ones that don’t (e.g., “Bluetooth pairing attack” doesn’t apply if your device has no Bluetooth).
  3. Add device-specific scenarios the library doesn’t cover.

Walking through each letter — example for our pump

S — Spoofing (CAW = 1.2)

Ask: how could an attacker pretend to be someone or something legitimate?

Network-attacker scenarios:

Physical / insider-attacker scenarios:

Detection methods (Framework V):

Compensating controls (Framework I):

CCD coverage assessment: Partial — physical fingerprinting and badge correlation cover most scenarios; vendor impersonation requires improved staff training.

T — Tampering (CAW = 1.1)

Ask: how could an attacker modify the device, its firmware, its configuration, or data in transit?

This is often the highest-stakes category for legacy medical devices because the consequence of successful tampering is direct patient harm (wrong dose, wrong therapy delivery, wrong sensor calibration).

Look especially at:

For our pump:

Network-attacker scenarios:

Physical / insider-attacker scenarios:

Detection methods (Framework V):

Compensating controls (Framework I):

CCD coverage assessment: Strong if MFA shim deployed; weak without it because the documented service-mode PIN is a low-bar attacker capability.

R — Repudiation (CAW = 0.9)

Ask: could an action on the device be later disowned by the actor who performed it?

For our pump: no per-user authentication means all configuration changes appear as “service-mode user”, impossible to attribute. This is a repudiation-by-default condition.

Network-attacker scenarios:

Physical / insider-attacker scenarios:

Detection methods:

Compensating controls:

CCD coverage assessment: Solved if MFA shim deployed; unsolvable without compensating control because the device cannot be modified to support per-user auth.

I — Information disclosure (CAW = 1.2)

Ask: could data flowing through, stored on, or displayed by the device be observed by an unauthorized party?

For our pump:

Network-attacker scenarios:

Physical / insider-attacker scenarios:

Detection methods:

Compensating controls:

D — Denial of Service (CAW = 1.5 — highest weight)

CAW is 1.5 because in clinical settings, DoS is not an inconvenience — it can be patient harm. A pump that stops infusing during chemotherapy is a harm event.

Network-attacker scenarios:

Physical / insider-attacker scenarios:

Detection methods:

Compensating controls:

E — Elevation of privilege (CAW = 1.1)

Ask: could an attacker gain higher-privileged access than they started with?

Network-attacker scenarios:

Physical / insider-attacker scenarios:

Detection methods:

Compensating controls:


Step 4 — Set Clinical Availability Weights (1 minute)

CAW values reflect the clinical consequence multiplier for each category. The defaults are:

Category CAW Why
Denial of Service 1.5 DoS = potential patient harm in clinical settings
Spoofing 1.2 Wrong-attribution scenarios drive wrong-care decisions
Information disclosure 1.2 PHI exposure is a HIPAA reportable event
Tampering 1.1 High consequence but moderated by detection feasibility
Elevation of privilege 1.1 Sets up other categories rather than being the harm event itself
Repudiation 0.9 Lowest because it amplifies other harms rather than causing them

Use the defaults unless you have a device-specific reason to adjust. Examples of valid adjustments:

Document the rationale next to the weight if you change a default.


Step 5 — Validate your work (3 minutes)

Walk through this checklist before declaring the threat model complete:

Compare against the worked example: examples/infusion-pump.md for Archetype 2, or examples/pacs-workstation.md for Archetype 1. Your completed model should resemble these in structure and density. If your file is much shorter than the worked examples, you’ve probably under-specified scenarios.


Step 6 — Where the output goes

A completed STRIDE-HC threat model is not the deliverable — it’s an input to four downstream artifacts:

Downstream artifact What the STRIDE-HC model provides Where to send it
FDA §524B premarket submission (vendors) The threat-model deliverable required by 2023 cybersecurity guidance Attach as Appendix to your premarket submission
ISO 14971 risk management file (vendors + HDOs) Hazard identification (cl.5) and risk estimation (cl.6) records Attach as the security-hazard section of your risk management file
AAMI TIR57 security risk records The security risk identification activities required by TIR57 Reference by ID in your TIR57 records
CJR documents (../cjr-templates/) The threat each compensating control is justified against Each STRIDE-HC scenario you cannot fully mitigate produces a CJR; cross-reference the STRIDE-HC scenario ID inside the CJR
MDRS calculator (../mdrs-calculator/) Empirical input to the CCD (Compensating-Control Deficit) score When you re-run MDRS after deploying controls, your CCD score should improve to reflect the new coverage

Review cadence for the threat model itself:

Sign-offs typically required:


Common pitfalls

Pitfall Why it happens Fix
All scenarios are network-attacker Authors come from IT security, not clinical engineering Force yourself to list at least one physical scenario per category before moving on
Compensating controls are aspirational “We could deploy 802.1X” without a plan Each control gets a CJR; the CJR forces you to confirm feasibility, cost, and timeline
CCD coverage marked “Strong” everywhere Wishful thinking Run the test harness (../test-harness/) to get empirical CCD evidence
Same scenario copied under multiple categories Confusion about which STRIDE letter applies Pick the letter that captures the attacker’s primary intent; cross-reference if needed but don’t duplicate
CAW values changed without rationale Following a non-clinical security template Use defaults; deviations must have inline rationale