Walkthrough — How the five artifacts fit together

This walkthrough demonstrates the integration claim of the companion paper: the five artifacts in this repository are designed as a connected workflow, not five standalone tools. Each step’s output is the next step’s input.

We use a hypothetical legacy infusion pump as the running example, because it’s the only device that appears across all five artifacts — making the chain visible and verifiable end-to-end. Substitute your own device; the flow is identical.

The chain in one picture

Inventory device     MDRS              STRIDE-HC          CJR              Test harness      MFA shim
─────────────        ─────             ─────────          ───              ────────────      ────────
Fact about device → Risk tier      →  Threat list      → Justified     → Empirical       → Deployed
                                      per-attacker       compensating    evidence           control
                                      per-category       control(s)
                                                                                              ↑
                                                                                              │
                                                          Pattern C is ONE specific control ──┘
                                                          (the other controls come from
                                                           paper §3 Compensating Controls
                                                           Playbook)

The scenario

You’re the security lead at a hospital. Clinical engineering surfaces 240 ExampleMed volumetric infusion pumps, model v3.2, running VxWorks 6.9 firmware. The vendor confirms: no patches coming, no MFA on the management port, hardcoded service-mode password documented in the service manual. ICU and oncology run these continuously on patients.

What do you do? Five artifacts, five steps.


Step 1 — Triage: how urgent is this? → MDRS Calculator

Open the calculator:

The calculator runs in guided mode by default — answer five plain-English multiple-choice questions and it assigns the appropriate score. Or click Direct entry to use sliders if you already know the scores. For our infusion pump:

Dimension What it asks Answer for this pump Score
CIS — Clinical Impact Severity Could a compromise harm a patient? Yes — wrong dose can kill 9.0
ES — Exploitability How attackable is it? Hardcoded password + cleartext mgmt protocol 7.5
DCI — Data Criticality & Integrity What data does it expose/control? Live infusion params + PHI 8.0
NEF — Network Exposure Factor How reachable is it? Sits on shared clinical VLAN, no segmentation 8.0
CCD — Compensating-Control Deficit How weak are existing mitigations? Some IDS, no segmentation, no PAM 7.0

Composite = 8.175 → CRITICAL. The irreversibility floor (CIS ≥ 9 → minimum HIGH) confirms this is not a borderline call. The “What to do” panel tells you: CISO + CMO + clinical engineering director within 24 hours, quarterly review.

Click Copy summary or Export JSON in the calculator to capture the result for the next step.

Output of Step 1: a tier (CRITICAL) and an executive-defensible composite number (8.175), plus next-step ownership and review cadence.


Step 2 — Model the threats: what specifically could go wrong? → STRIDE-HC Templates

Open: stride-hc-templates/examples/infusion-pump.md — the worked example for this exact device archetype (Archetype 2 = embedded RTOS legacy).

The template walks through all six STRIDE-HC categories — Spoofing, Tampering, Repudiation, Information disclosure, DoS, Elevation of privilege — and for each one asks two questions:

  1. What’s the network-attacker scenario?
  2. What’s the physical/insider-attacker scenario?

The second question is the key STRIDE-HC modification: the physical attacker is treated as a co-equal threat surface, not an afterthought.

For our pump, here are two threats the template surfaces (excerpting from the worked example):

T — Tampering (CAW = 1.1)
  Network:  Unauthorised firmware push from compromised vendor channel
  Physical: Firmware injection via USB or service port           ← high-priority
  Detection:  File integrity monitoring (network-adjacent)
  Compensating controls:
    - USB port disablement
    - Inline MFA shim (Pattern C) at service port                ← solution candidate
    - Tamper-evident sealing

If you’re not starting from the worked example, the scenario library at stride-hc-templates/scenario-library/archetype-2-scenarios.md (or archetype-1-scenarios.md) gives you a pre-filled menu so you don’t stare at a blank page.

For a first-time tutorial that builds a STRIDE-HC threat model from blank to completed, see stride-hc-templates/WALKTHROUGH.md.

Output of Step 2: a list of specific threat scenarios + a list of candidate compensating controls per threat.


Step 3 — Pick a control: which compensating control do you deploy?

For our pump, the highest-priority threat is physical-attacker firmware injection at the service port — because (a) the device is CRITICAL from Step 1, (b) the service port is unauthenticated, and (c) the consequence is uncontrollable infusion behaviour.

The STRIDE-HC template surfaced three candidate controls:

  1. USB port disablement — rejected: breaks the vendor’s authorized service workflow
  2. Inline MFA shim at service port (Pattern C)chosen
  3. Tamper-evident sealing — retained as a complementary control

You pick #2 because USB disablement makes scheduled service impossible, and sealing gives detection but not prevention. Sealing is kept as a layered control.

This decision needs to be documented. That’s Step 4.


Step 4 — Document the decision: prove it’s defensible → CJR Templates

Open: cjr-templates/examples/cjr-no-mfa-infusion-pump.md — the worked Control Justification Record for this exact decision (deploying an MFA shim because the device can’t run MFA natively).

A CJR is a regulator-readable, audit-readable document that captures:

  1. The device — with explicit cross-link to the MDRS tier from Step 1 and the STRIDE-HC model from Step 2
  2. The standard control that cannot be applied — here, native MFA on the management interface
  3. The constraint preventing the standard control — vendor will not re-cert firmware with MFA; FDA 510(k) burden too high
  4. The threat being addressed — mapped to specific STRIDE-HC scenarios by ID (Step 2’s output is referenced, not duplicated)
  5. The compensating control deployed — MFA shim Pattern C
  6. Standards mapping — HIPAA §164.308, ISO 14971 cl.7, AAMI TIR57, FDA §524B
  7. Sign-offs — CISO, clinical engineering, regulatory affairs, CMO
  8. Review cadence — matches the tier from MDRS (quarterly for CRITICAL)

Notice the explicit cross-references in the example file:

| Linked STRIDE-HC threat model | stride-hc-templates/examples/infusion-pump.md |
| Current MDRS score and tier   | 8.175 → CRITICAL                              |

The CJR is the evidence artifact you show an FDA auditor, a HIPAA Security Rule reviewer, or your board. It justifies why you chose the compensating control over the standard one and why that’s appropriate for this device’s tier.

Output of Step 4: a signed, dated, audit-ready document.


Step 5 — Empirically prove the control works → Test Harness

Open: test-harness/README.md and test-harness/METHODOLOGY.md.

A CJR is paperwork. Auditors increasingly ask: “You claim this compensating control works. Show me data.” That’s what the test harness is for.

It’s a Dockerised multi-container environment:

docker compose up

┌─────────────────┐       ┌──────────────────────┐       ┌──────────────┐
│  Attacker       │       │ Compensating control │       │ Pump emulator│
│  (5 scenarios,  │  ───▶ │ (IPS / segmentation /│  ───▶ │ (VxWorks-like│
│   mapped to     │       │  PAM)                │       │  responses)  │
│   STRIDE)       │       └──────────────────────┘       └──────────────┘
│  - ARP poison   │                  │
│  - firmware inj │                  ▼
│  - cleartext HL7│           Results CSV (with-control vs without-control)
│  - DoS flood    │
│  - default cred │
└─────────────────┘

The five attack scenarios in test-harness/attacker/scenarios/ map 1-to-1 to the STRIDE-HC letters from Step 2. Each runs twice — once with the compensating control enabled, once without — and writes to test-harness/results/sample-results.csv.

Output of Step 5: a CSV with empirical pass/fail evidence per scenario, which you attach to the CJR as Section 6 evidence (“control effectiveness validated by harness run XYZ on YYYY-MM-DD”).


Step 6 — Deploy the actual hardware control → MFA Shim (Pattern C)

Open: mfa-shim/README.md — reference design, hardware BOM, FDA considerations.

The MFA shim is the only artifact in the suite that’s a specific compensating control, not a framework. It’s the Pattern C from the paper: an inline device that sits between the technician’s laptop and the medical device’s service port, intercepts the connection, requires the technician to authenticate via TOTP + session binding, then proxies the cleartext service-mode session while recording every byte for forensic accountability.

The reference design at mfa-shim/prototype/ is the Python implementation with three components:

File Role
totp_gate.py The authentication gate (TOTP, window tolerance, lockout, per-user rate limiting)
session_recorder.py Forensic session recording with SIEM forwarding
tamper_detector.py Tamper sensors + emergency stop

37 unit tests prove each component behaves correctly under adversarial conditions (replay, lockout bypass, sensor failures, concurrent authentication, etc.). Run them with:

cd mfa-shim/prototype
python -m venv .venv && source .venv/bin/activate  # Windows: .venv\Scripts\activate
pip install -r requirements.txt
python -m pytest tests/ -v

mfa-shim/FDA-CONSIDERATIONS.md addresses the regulatory question of inline-device modification to a cleared medical device — read this before any clinical deployment.

Output of Step 6: a deployable reference design with FDA-considered architecture, ready for adaptation by a vendor or HDO building the production version.


How the artifacts cross-reference each other

You can verify the integration claim by following the explicit cross-links inside the worked examples:

This is the integration the paper claims: not five tools sharing a directory, but five tools sharing a workflow.

Where to go next

If you want to… Read
Build a STRIDE-HC threat model from scratch stride-hc-templates/WALKTHROUGH.md
Write a CJR for a new compensating control cjr-templates/WALKTHROUGH.md — 30-minute first-author tutorial
Run the test harness test-harness/README.md
Understand the MDRS methodology mdrs-calculator/docs/methodology.md
Adapt the MFA shim for production hardware mfa-shim/ARCHITECTURE.md + mfa-shim/FDA-CONSIDERATIONS.md
Cite this work README.md — Citing this work section